Abstract
In liquid chromatography-mass spectrometry (LC-MS)-based proteomics, many precursors elute from the column simultaneously. In data-dependent analyses, these precursors are fragmented one at a time, whereas the others are discarded entirely. Here we employ trapped ion mobility spectrometry (TIMS) on an orthogonal quadrupole time-of-flight (QTOF) mass spectrometer to remove this limitation. In TIMS, all precursor ions are accumulated in parallel and released sequentially as a function of their ion mobility. Instead of selecting a single precursor mass with the quadrupole mass filter, we here implement synchronized scans in which the quadrupole is mass positioned with sub-millisecond switching times at the m/z values of appropriate precursors, such as those derived from a topN precursor list. We demonstrate serial selection and fragmentation of multiple precursors in single 50 ms TIMS scans. Parallel accumulation-serial fragmentation (PASEF) enables hundreds of MS/MS events per second at full sensitivity. Modeling the effect of such synchronized scans for shotgun proteomics, we estimate that about a 10-fold gain in sequencing speed should be achievable by PASEF without a decrease in sensitivity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.