Abstract

We present Monte-Carlo simulations for the focusing design of a novel cold-neutron triple-axis spectrometer to be installed at the end position of the cold guide NL-1 of the research reactor FRM-II in Munich, Germany. Our simulations are of general relevance for the design of triple-axis spectrometers at end positions of neutron guides. Using the McStas program code we performed ray trajectories to compare parabolic and elliptic focusing concepts. In addition the design of the monochromator was optimized concerning crystal size and mosaic spread. The parabolic focusing concept is superior to the elliptic alternative in view of the neutron intensity distribution as a function of energy and divergence. In particular, the elliptical configuration leads to an inhomogeneous divergence distribution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.