Abstract

A compact Raman system constructed by a parabolic sample cell and an imaging spectrograph, which has good capability for enhancing a Raman signal and compressing a continuous background, has been put forward. In the Raman spectra of ambient air acquired by this system, the signal level of N2 was enhanced up to 14 times compared with free space, and the related signal to background ratio was increased nearly to 96. With an integration time of 10 s, the rotational fine structure of O2 and N2 were clearly recognized. Besides, a standard analytic gas mixture consisting of H2, CO2, and CO was also tested, and the 3σ LODs of 68 ppm for H2, 54 ppm for CO2 and 116 ppm for CO were obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.