Abstract

A fast methodology for the assay of parabens in drug and cosmetic preparations has been presented. The procedure developed is based on paper spray tandem mass spectrometry and isotope dilution approach. For each investigated paraben, the corresponding labeled standard has been used in order to improve the accuracy and reproducibility of the analyses. The MS experiments have been performed under MRM conditions, monitoring the transitions [M-H]- → m/z 92 and [M-H]- → m/z 98, respectively, for each analyte and the corresponding labeled internal standard. The quantitative assay has been performed using a calibration curve built from 2 to 15 mg/L. The method accuracy, in all case near 100%, was evaluated using fortified samples at two concentration levels, which are representative of the lower and the higher portion of calibration curve. The good values of LOQ, LOD, and reproducibility confirm the consistency of the developed approach.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.