Abstract
A simple paper-based sensor with electrochemical detection was first developed for rapid, selective, and sensitive determination of acetylcholinesterase (AChE). The screen-printed graphene electrode was used as working electrode providing sensitivity for the sensor. The amperometric detection of AChE is based on the determination of thiocholine (TCh) produced from hydrolysis of acetylthiocholine chloride (ATCh) by AChE. To detect TCh, the ATCh immobilized sheet was stacked onto the detection sheet using double adhesive tape, then samples of AChE were dropped on the back side of an ATCh immobilized sheet with only 1min of incubation time. To avoid interference, glutathione (GSH), the potential of 0.5V vs. Ag/AgCl was applied onto a graphene electrode and the current, which depends on AChE concentration, was measured. Under optimized conditions, the limit of detection (LOD) from the experiment of AChE determination was 0.1U/mL with AChE concentration in range of 0.1–15U/mL. The data correlated well with the data obtained using spectrophotometric method. The developed sensor had been successfully applied to detect AChE in blood samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.