Abstract

The performance assessment and beam characteristics of the Therapax SXT 150 unit, which encompass both low and medium-energy beams, were evaluated. Dose determination was carried out by implementing the International Atomic Energy Agency (IAEA) TRS-398 protocol and measuring all the dosimetric parameters in order to have a solid, consistent and reliable data set for the unit. Mechanical movements, interlocks and applicator characteristics agreed with specifications. The timer exhibited good accuracy and linearity. The output was very stable, with good repeatability, long-term reproducibility and no dependence on tube head orientation. The measured dosimetric parameters included beam first and second half-value layers (HVLs), absorbed dose rate to water under reference conditions, central axis depth dose distributions, output factors and beam profiles. Measured first HVLs agreed with comparable published data, but the homogeneity coefficients were low in comparison with typical values found in the literature. The timer error was significant for all filters and should be taken into consideration for the absorbed dose rate determination under reference conditions as well as for the calculation of treatment times. Percentage depth-dose (PDD) measurements are strongly recommended for each filter-applicator combination. The output factor definition of the IAEA TRS-398 protocol for medium-energy X-ray qualities involves the use of data that is difficult to measure. Beam profiles had small penumbras and good symmetry and flatness except for the lowest energy beam, for which a heel effect was observed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call