Abstract

The fast control of the beam spill extracted from a synchrotron is a key function for the spot scanning irradiation in cancer therapy application. The authors propose an extraction method which uses the quadruple field of fast response, as well as the RF-knockout. A Panofsky magnet was developed as a quadruple magnet, with a frequency response of around 10 kHz. The Panofsky magnet has a rectangular beam aperture and plate coils attached to the pole face. A model magnet has been manufactured with ferrite, and static and dynamic magnetic fields were measured. From the measurement we observed that the effects of eddy current in the plate coils were large and the uniformity of the magnetic field gradient in the beam aperture was worse than ±5% with a plate thickness of 0.02 cm and a frequency of current of 10 kHz. For the future, in a detailed design the eddy current effects have to be taken into account.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.