Abstract

A Carpentier-Edwards pericardial (CEP) bioprosthesis was explanted from an 81-year-old woman due to nonstructural dysfunction 9 years after mitral valve replacement. The nonstructural dysfunction produced severe regurgitation in the mitral position. During the surgery, excessive pannus overgrowth was seen on the left ventricular side of the CEP bioprosthesis. Pannus overgrowth was prominent on one leaflet. That leaflet was stiff and shortened due to the excessive overgrowth of pannus. In this patient, the distortion of one leaflet was the main reason for transvalvular leakage of the CEP bioprosthesis in the mitral position. A new CEP bioprosthesis was implanted in the mitral position. Pathological analysis revealed fibrotic pannus with a small amount of cellular material over the leaflets of the resected CEP valve. This change was marked on the distorted leaflet.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.