Abstract

To compare pancreatic virtual unenhanced (VUE) and true unenhanced (TUE) images and to calculate the potential dose reduction by omitting the conventional unenhanced scan. Fifty-one patients with known or suspected pancreatic masses underwent contrast-enhanced computed-tomography (CT) during unenhanced and portal venous phases acquired in single-energy (SE) mode, and pancreatic parenchymal phase acquired in dual-energy (DE) mode. The image quality (IQ) and image noise (IN) of TUE and VUE images were evaluated. The effective dose of a combined DE/SE dual-phase protocol was compared with that of a theoretical standard SE triple-phase protocol. The results were tested for statistical significance using the Cohen's k, the Wilcoxon's signed rank test, and the paired t-test; p-values of less than 0.05 were considered significant. Mean TUE and VUE IQ were 1.5±0.6 and 1.6±0.6 (k=0.891), with no significant difference (p>0.05). Mean TUE and VUE IN were 12.3±1.6 and 10.3±1.5HU, and resulted significantly different (p<0.001). Mean effective doses for a combined DE/SE dual-phase protocol and SE triple-phase protocol were 8.9±2.4mSv (range 4.8-16.2mSv) and 12.1±3.1mSv (range 6.4-21.1mSv). The calculated mean dose reduction achievable by omitting the unenhanced scan was 26.7±9.7% (range 10-46.1; p<0.001). VUE images are feasible for pancreatic abdominal CT. A combined DE/SE dual-phase protocol permits a significant reduction in dose exposure to patients.