Abstract

Transient elastography (TE), recommended by the WHO, is an established method for characterizing liver fibrosis via liver stiffness measurement (LSM). However, technical barriers remain towards point-of-care application, as conventional TE requires wired connections, possesses a bulky size, and lacks adequate imaging guidance for precise liver localization. In this work, we report the design, phantom validation, and clinical evaluation of a palm-sized TE system that enables simultaneous B-mode imaging and LSM. The performance of this system was validated experimentally using tissue-equivalent reference phantoms (1.45-75 kPa). Comparative studies against other liver elastography techniques, including conventional TE and two-dimensional shear wave elastography (2D-SWE), were performed to evaluate its reliability and validity in adults with various chronic liver diseases. Intra- and inter-operator reliability of LSM were established by an elastography expert and a novice. A good agreement was observed between the Young's modulus reported by the phantom manufacturer and this system (bias: 1.1-8.6%). Among 121 patients, liver stiffness measured by this system and conventional TE were highly correlated (r = 0.975) and strongly agreed with each other (mean difference: -0.77 kPa). Inter-correlation of this system with conventional TE and 2D-SWE was observed. Excellent-to-good operator reliability was demonstrated in 60 patients (ICCs: 0.824-0.913). We demonstrated the feasibility of employing a fully integrated phased array probe for reliable and valid LSM, guided by real-time B-mode imaging of liver anatomy. This system represents the first technical advancement toward point-of-care liver fibrosis assessment. Its small footprint, along with B-mode guidance capability, improves examination efficiency and scales up screening for liver fibrosis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call