Abstract

To evaluate the safety and long-term effectiveness of a new treatment for snoring using permanent palatal implants. Prospective study with 35 patients referred to the clinic for the treatment of snoring. The patients were followed for 1 year, and a sleep analysis was performed and repeated at 360 days. During the procedure, 3 braided polyethylene terephthalate implants were placed in the palate of each patient. The average duration for nonnarcotic analgesics was 1.3 days. No adverse events were observed. The mean snoring intensity was reduced from 7.1 at baseline to 4.8 at 1 year (P < 0.001). Bed partner satisfaction rate was very high at 70.6% after 1 year. Six patients (17.8%) experienced a total of 9 partial extrusions. The palatal implant system is safe, and the effectiveness in treating snoring is comparable to other surgical methods. Advantages are virtually no tissue damage and minimal patient discomfort as a result of one standardized, low-cost procedure. C-4.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.