Abstract
To test wireless resonance circuits (RC) to be used as fiducial marker of endovascular catheters during MR-guided interventions. Current markers loose their signal enhancement for certain catheter orientations. The purpose of this study was to test a marker setup which overcomes this orientation problem. The markers were constructed from a pair of two RCs. The RCs were individually tuned and the coil axes were oriented perpendicular to each other in order to decouple the two RCs. The markers were mounted on the tip of endovascular catheters and tested in vitro and in one porcine in vivo experiment. An intense MR signal at similar signal levels was noted at all catheter orientations. In the in vivo experiment the markers allowed for fast and reliable MR guidance of the catheters. A pair of two individually tuned and decoupled RCs is well suited for MR guidance of endovascular catheters.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.