Abstract

Infants born with complete atrioventricular block (CAVB) and fetal bradycardia are frequently born with low birth weight. Three low-birth-weight CAVB infants underwent temporary pacemaker implantation, followed by permanent single-chamber pacemaker implantation at median body weights of 1.7 and 3.2 kg, respectively. All infants caught up with their growth curves and had >3 years of estimated residual battery life. This two-stage strategy was successful in facilitating permanent pacemaker implantation in low-birth-weight babies. Placement of single-chamber pacemaker on the apex of the left ventricle appears to be associated with longer battery lifespan.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.