Abstract

One fundamental issue of the use of high-power microwave (HPM) sources is the health impact, in particular for patients who wear active implantable medical devices (AIMDs). This paper presents the effects of HPM ultrawideband (HPM/UWB) radiation on a pacemaker exposed to them, which is one aspect of the compatibility between HPM sources and AIMDs. The pacemaker was placed inside a phantom and exposed to a pulsed electromagnetic field with a rise time of the order of 100 ps and with an amplitude ranging between 350 kV/m and 1 MV/m. These electromagnetic field amplitudes are much higher than those used in classical scenarios involving HPM/UWB sources. Different configurations and programming modes of the pacemaker were tested. On the whole, the results of the experiments show the relative hardness and reliability of the pacemaker subjected to HPM/UWB radiation. A few nonpermanent disruptions occurred when the pulse repetition frequency and the exposure duration reached 100 Hz and 10 s, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.