Abstract

A conventional apparatus for determination of oxygen in organic compounds has been improved for application to organic fluorine compounds. A feature of the apparatus is the use of a pyrolysis tube made of glassy carbon instead of quartz, which eliminates effects due to hydrogen fluoride produced in pyrolysis of the sample. Ten analyses of dexamethasone with the apparatus gave a mean value of 20.44% for oxygen (theory, 20.38%), with a standard deviation of 0.16%. Oxygen in 9 organic fluorine compounds was accurately determined by using the apparatus, with an average error of +0.1%. One analysis by a gravimetric or a coulometric method took about 40 or 25 min, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.