Abstract
This study was aimed at comparing the oxidative degradation of commercial acetabular cups made of cross-linked polyethylene (XLPE) and conventional ultrahigh molecular weight polyethylene (UHMWPE). After testing against deliberately scratched CoCrMo femoral heads in a hip joint simulator, the cups, microtomed parallel to the articulating surface, were analyzed by IR spectroscopy. Due to the potential for artifacts caused by absorbed contaminants, the IR spectra were compared only after hexane extraction; actually, XLPE was found to absorb more serum than UHMWPE. The two sets of unworn acetabular cups showed different oxidation patterns with consequently different distributions of carbonyl species; unworn XLPE was characterized by lower contents of carbonyl species and hydrogen-bonded alcohols and higher contents of trans-vinylene species than unworn UHMWPE. Upon simulator testing, UHMWPE showed more significant changes in oxidation indexes and distribution of carbonyl compounds than XLPE, confirming a better wear behavior for XLPE under the adopted testing conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.