Abstract

Tricuspid regurgitation (TR) is common and associated with significant mortality and morbidity. Because the effectiveness and safety of medical and surgical treatments are limited, there is a significant unmet need for the treatment of this disease. Therefore, there is a growing market for percutaneous devices that offer safer, less invasive, and more effective treatment options in this patient population. On February 13, 2024, the US Food and Drug Administration (FDA) convened a meeting of the Circulatory System Devices Panel to discuss the safety and effectiveness of the TriClip Transcatheter Valve Repair System (Abbott, Santa Clara, CA, USA). Several important points were discussed, including newly published data from the TRILUMINATE Pivotal study, the use of patient-oriented outcomes for device approval, and a discussion about training requirements and rollout plans when approving a breakthrough device. In this manuscript, we summarize the data presented by the sponsor and FDA and describe the deliberations and discussions during the meeting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.