Abstract

We have recently observed an increasing number of patients presenting very high serum levels of 25-hydroxyvitamin D [25(OH)D] (> 150 ng/mL), which, in all cases, had been measured with the IDS EIA kit adapted on different "open" automated platforms. We performed a comparison between the IDS EIA kit adapted on two different "open"automated platforms and the DiaSorin RIA. We found a systematic bias (higher levels with the IDS EIA kit) for concentrations more than 50-60 ng/mL that was less obvious when the IDS EIA was used in its manual procedure. We thus suggest to use the IDS EIA kit in its manual procedure rather than to adapt it on an automated platform, and to interpret cautiously a 25(OH)D greater than 100 ng/mL with this kit.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.