Abstract
Patients with end-stage cardiomyopathy and congestive heart failure are increasingly undergoing implantation with left ventricular assist devices (LVADs). In addition, implantable cardioverter-defibrillator (ICD) therapy has been proven to be an important part of the treatment for cardiomyopathy/congestive heart failure. Previous reports have noted a potential and dramatic electromagnetic interference from LVADs on ICDs that cause impaired telemetry communication between the ICD and ICD programmer. Such interference has necessitated explantation and generator replacement in order to resume communication between the ICD and programmer. We report two patients with advanced congestive heart failure and ICD programming impairment caused by a HeartMate II LVAD (Thoratec Corporation, Pleasanton, CA, USA) that was overcome by placing aluminum shielding around the ICD programmer wand and steel shielding around the extension cable during ICD interrogation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.