Abstract
The aim of this study was to compare oval and round window vibroplasty. Eighteen (18) patients implanted with Vibrant Soundbridge (VSB) were enrolled. Two groups were formed depending on FMT placement: on round window in ten cases (RW group) and on oval window in eight (OW group). Pre and postoperative audiological tests were performed both under headphones and free-field settings, VSB on and off. One (1) RW patient experienced sudden hearing loss at the operated side after 4 months from surgery and was excluded from the analysis. Both groups showed good hearing results. Significant differences were measured at free-field pure-tone test with VSB on at 0.5 kHz (RW better than OW, p = 0.026) and 4 kHz (OW better than RW, p = 0.043). Both techniques share similar good results and are considered safe. However, we had one failure with deep and sudden hearing threshold worsening after some months of good results. From a surgical point of view OW vibroplasty is easier and safer to perform, when the stapes suprastructure is absent, as it does not require any drilling and should be preferred in such cases. More reports are needed to explain if RW vibroplasty is risky in a mid to long term.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.