Abstract

This study investigates the output from blocked treatment fields with 6 MV and 18 MV photons. Square fields, defined by collimators, were blocked down to smaller symmetrical square, rectangular and asymmetrical rectangular fields using blocks on a tray beneath the collimators. Up to 7.5% difference was found between the output measured for the blocked field and the output calculated using a collimator output factor for the surface field size. A method is described by which output factors for the blocked fields can be measured and used in practice. Three methods of approximating these output factors are also presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.