Abstract

Outcomes of device closure of large and small secundum atrial septal defects (ASDs) as related to rim anatomy with the Amplatzer atrial septal occluder were compared. Rim adequacy (> or = 5mm) of the anterior, inferior, posterior, and superior rims was determined using transesophageal echocardiography. Balloon-stretched defect size defined patients into two groups: group 1, < or = 25 mm (n = 138); group 2, > 25 mm (n = 34). Rim deficiency (n = 62) was more frequent in group 2 compared to group 1 (50% vs. 33%; P = 0.07), especially inferior rim deficiency (35% vs. 2%; P = 0.005). Device deployment was successful in group 1 and group 2 (100% vs. 91%; P = 0.007). Unsuccessful deployment was associated with an ASD of > 25 mm (P = 0.007) and inferior rim deficiency (P = 0.001). At first follow-up (54 +/- 16 days), right ventricular systolic pressure had improved in both groups (P < 0.001). Closure of a large ASD associated with a lack of support in the inferior rim may warrant alternative strategies to position the device successfully.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.