Abstract

During the period from April 4, 1988, to April 20, 1988, nine pyrogenic reactions and five gram-negative bacteremias occurred in 11 patients undergoing dialysis. All pyrogenic reactions and gram-negative bacteremias occurred among patients in whom a reprocessed dialyzer was used. The rate of pyrogenic reactions or bacteremias per 100 sessions using a reprocessed dialyzer was higher than in sessions during which a new dialyzer was used (4.5 vs. 0; p = 0.03). Dialyzers were manually reprocessed with 2.5% Renalin germicide. The Renalin concentrations varied widely in 12 dialyzers stored after manual reprocessing during the epidemic period (0.9-4.2%); the median endotoxin concentrations varied from 0 to 246 ng/ml and were higher in dialyzers with Renalin concentrations less than or equal to 1.0% than in dialyzers with higher concentrations (p = 0.01). Experiments using a dilution technique described by a technician resulted in Renalin concentrations ranging from 1.4% at the surface to 3.5% at the bottom of the preparation container. These findings suggest that failure to adequately admix Renalin during dilution may be associated with low levels of disinfectant, high levels of bacteria and endotoxins in dialyzers, and outbreaks of pyrogenic reactions and gram-negative bacteremias in dialysis patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.