Abstract

AbstractBackgroundElectron out‐of‐field scatter is generally not given importance mainly in electron fields. However, this is important when applicator down and boost treatments are given usually at an angle from the central axis. The electron scatter dose is found to be far away from the central axis which could be easily ignored.PurposeThis study aims to investigate the out‐of‐field radiation doses from electron applicators and their effects on clinical treatment. By identifying the parameters that contribute to out‐of‐field doses and to explore potential strategies for reducing these doses in order to improve patient outcomes from modern machines.MethodsMeasurements were performed in water phantom using electron diode for modern Elekta and Varian machines. Dose profiles were acquired at surface and dmax with 0° and 90° collimation angle. Various gantry angles were also studied for some data with IC Profiler. The profiles were normalized with respect to the central axis dose.ResultsThe scatter dose peaks were found at a distance between 11 and 28 cm from the central axis on all machines. However, the peak shifts to 15 cm at 90° collimator when beam is tilted. The position and intensity of the dose varies with depth, collimator, and gantry angles for both Elekta and Varian machines. Due to clearance issues more gantry angles were studied for Elekta applicator compared to Varian. In general, Varian TrueBeam has a lower scatter that Elekta Infinity. The 90° collimator angle has a higher scatter compared to zero degree for both machines.ConclusionsThere are clinically significant peripheral doses around 3% of the central axis dose from the electron applicator. Elekta has a slightly higher scatter (3%) than Varian (2%) that peaks at 25 cm which is clinically important but often overlooked.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call