Abstract
Late failure of total knee arthroplasties usually results from ultrahigh molecular weight polyethylene wear or implant loosening. Early failure from osteolysis is uncommon. However, we treated a patient with a failed total knee arthroplasty from osteolysis that developed 2 years postoperatively. The failure was associated with tibial component debonding from the cement mantle with abundant cement and metal debris. Although there was some third-body debris in the ultrahigh molecular weight polyethylene insert surface, the insert wear was not extensive. Although abundant cement and metal debris were found in the periarticular soft tissues, no ultrahigh molecular weight polyethylene was seen in histologic specimens under polarized light. The osteolysis seems to have been caused primarily by debris generated from debonding and torsional motion at the tibial baseplate-cement interface rather than the bearing surface. Although this failure mechanism has been well recognized in cemented total hip arthroplasties, it has not been reported to be a substantial cause of failure in total knee arthroplasties.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.