Abstract

Impedance, or oscillometry, measurements of the respiratory system can generate information about the function of the respiratory system not possible with traditional spirometry. There are currently several instruments on the market using different perturbations. We have compared a new respiratory oscillometry instrument, the tremoflo, with Impulse Oscillometry (IOS). Patients with a physician’s diagnosis of chronic obstructive lung disease (COPD) and healthy subjects were recruited. They underwent assessment of respiratory function with oscillometry using the IOS and tremoflo devices and the resulting impedance data from the two methods were compared. The two devices were also tested against a reference respiratory phantom with variable resistances. Whereas both devices detected impairments in the patients’ lung function commensurate with small airways pathology, the tremoflo appeared to be more sensitive than the IOS. We found systematic differences between the two instruments especially for reactance measurements where the area over the reactance curve (AX) was significantly lower with the IOS compared with the tremoflo (p < 0.001). Moreover, the agreement between the two devices was reduced with increasing severity of the disease as determined with a Bland-Altman test. Testing both instruments against a respiratory phantom unit confirmed that the resistance measured by the tremoflo compares closely with the known resistance of test loads, whereas the IOS’ resistance correlated with a test load of 0.19, kPa.s.L−1 at higher loads it deviated significantly from the known resistance (p < 0.0028). We conclude that the absolute values measured with the two devices may not be directly comparable and suggest that differences in the calibration procedures might account for the differences.

Highlights

  • The diagnosis of chronic obstructive pulmonary disease depends to a large degree on pulmonary function tests and the Global Initiative on Chronic Obstructive Lung Disease (GOLD) standard[1] is based on the assessment of forced exhalation maneuvers by the patient

  • Subjects with COPD of both sexes were recruited from patients attending the Dept. of Respiratory Medicine and referred for pulmonary function tests at the Department of Clinical Physiology, Skåne University Hospital, Malmö

  • The patients were all diagnosed with various degrees of COPD according to the GOLD standard[27]

Read more

Summary

Introduction

The diagnosis of chronic obstructive pulmonary disease depends to a large degree on pulmonary function tests and the Global Initiative on Chronic Obstructive Lung Disease (GOLD) standard[1] is based on the assessment of forced exhalation maneuvers by the patient. In subjects with COPD, the main findings in oscillometry are an increase in Resistance and a reduction of Reactance at the lower frequencies, and an increase of the area under the curve of Reactance[11] While both devices can detect alterations in lung function, we wanted to elucidate if the fundamental differences in the perturbation signal (single pulse v.s. a broadband sine-wave composite signal) might yield different results with the two systems. To address this question, we contrasted measurements of input impedance obtained with IOS and tremoflo. We constructed a lung phantom with which we made comparisons at different impedances with the two instruments

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call