Abstract

In this study, the effects of surface smoothness on fatigue characteristics of Ti-50.4at%Ni shape memory alloy (which is used as self-expanding stent) wire are investigated. The Ti-Ni specimen is a wire 1mm in diameter. The surface smoothness is varied by the physically polishing and electro-polishing. The surface and fractured surface are observed by scanning electron microscope (SEM). The surface of electro-polished specimen is smoother than the other specimen. The fatigue characteristics of specimen is investigated by the rotary bending tests in the water kept 310K. The applied vending strain (ε_<ap>) is varied from 1 to 6% in this study. When the ε_<ap>, is above 3%, fatigue characteristic of electro-polished specimen is better than the other specimen, while ε_<ap> is below 3%, physically polished specimen also become good as well as fatigue characteristic of electro-polished specimen. Furthermore, the fracture origins of all specimen are found on the surface of each specimen. In consequence, fatigue characteristics are improved with increasing of the surface smoothness.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.