Abstract
Backscattering of an incident ultrasound beam by blood gives rise to a Doppler spectrum whose characteristics are affected not only by the velocity distribution, but also by certain basic aspects of the RBC's behavior. Starting from fundamental assumptions, an explicit expression is derived for the Doppler spectrum in terms of the variance in the scatterer number density and the backscattering cross section. This shows that the Doppler power at a given frequency is weighted by the backscattering cross section and the manner in which the red blood cells (RBC's) are packed (packing factor). Since spatial variations in the flow field can result in changes in the variance and backscattering cross section, the mean Doppler frequency will not necessarily be proportional to the mean flow through the sample volume. Experimental results for two different flow fields are used to illustrate these effects. The implications of these findings are discussed in relation to volumetric flow estimation and the power mode display used in some color Doppler flow imaging systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.