Abstract

ObjectiveTemporally extended signal space separation (tSSS) is a powerful method for artifact suppression in magnetoencephalography (MEG). Because tSSS first separates MEG signals coming from inside and outside a certain sphere, definition of the sphere origin is important. For this study, we explored the influence of origin choice on tSSS performance in spontaneous and evoked activity from epilepsy patients. MethodsInterictal epileptiform discharges (IEDs) and somatosensory evoked fields (SEFs) were processed with two tSSSs: one with the default origin of (0, 0, 40 mm) in the head coordinate, and the other with an individual origin estimated using each patient’s anatomical magnetic resonance imaging (MRI). Equivalent current dipoles (ECDs) were calculated for the data. The ECD location and quality of estimation were compared across conditions. ResultsMEG data from 21 patients revealed marginal differences in ECD location, but the estimation quality inferred from goodness of fit (GOF) and confidence volume (CV) was better for the tSSS with individual origins. This choice affected IEDs more than it affected SEFs. ConclusionsIndividual sphere model resulted in better GOF and CV. SignificanceApplication of tSSS using an individual origin would be more desirable when available. This parameter might influence spontaneous activity more strongly.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call