Abstract

A new ultrasound-based device is proposed to non-invasively measure the orientation of the scapula in the standing position to consider this parameter for Total Shoulder Arthroplasty. The aim of this study was to assess the accuracy and reliability of this device. Accuracy was assessed by comparing measurements made with the ultrasound device to those acquired with a three-dimensional (3D) optical localization system (Northern Digital, Canada) on a dedicated mechanical phantom. Three users performed 10 measurements on three healthy volunteers with different body mass (BMI) indices to analyze the reliability of the device by measuring the intra and interobserver variabilities. The mean accuracy of the device was 0.9°± 0.7 (0.01-3.03), 1.3°± 0.8 (0.03-4.55), 1.9°± 1.5 (0.05-5.76), respectively, in the axial, coronal, and sagittal planes. The interobserver and intraobserver variabilities were excellent whatever the BMI and the users experience. The device is accurate and reliable enough for the measurement of the scapula orientation in the standing position.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.