Abstract

The organ and effective doses computed with a new flexible programme were compared in four X ray examinations with the doses determined with LiF TL detectors in an Alderson-Rando phantom. The correlation between the two methods was rather good (within 25%) for most results of the comparisons (75% of 45 results), but rather poor (discrepancy up to 140%) for the low and high density organs (lung, bone). An upward trend was also obtained for the doses to colon and some other organs. The reasons for the discrepancies were thought to be due to the approximate density and attenuation corrections of the computer programme for bone and lung tissues.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.