Abstract
Organ and effective doses during paediatric chest radiographic examination were investigated for various tube voltages between 60 and 110 kV at a constant milliampere-second value and focus-to-film distance by using an in-phantom dose measuring system and a Monte Carlo (MC) simulation software (PCXMC), where the former was composed of 32 photodiode dosemeters embedded in various tissue and organ sites within a 6-y-old child anthropomorphic phantom. Lung doses obtained ranged from 0.010 to 0.066 mGy and effective doses from 0.004 to 0.025 mSv, where these doses varied by a factor of 6 with the change in the tube voltage. Effective doses obtained using the MC simulation software agreed with those obtained using the dose measuring system within 23 %, revealing the usefulness of PCXMC software for evaluating effective doses. The present study would provide helpful dose data for the selection of technical parameters in paediatric chest radiography in Japan.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.