Abstract
Conversion coefficients are commonly used to estimate the radiation dose in projectional radiography from the dose-area product (DAP). This study investigates whether conversion coefficients can also be used for dosimetry in fluoroscopic procedures. Thermoluminescence dosimeters (TLD) were placed into an anthropomorphic phantom (Alderson Rando Phantom) to measure the dose absorbed by the gonads and uterus. A standardized enteroclysis in double contrast technique (Sellink/Herlinger) was simulated and the DAP was measured. Three examinations with 70, 80 and 90 kV were carried out with a fluoroscopy time of 15 minutes each. The relations between the DAP and the organ dose were analyzed and a linear regression was performed to determine the conversion coefficients [mSv/cGycm (2)]. The DAP and the organ dose measured by TLD were strongly correlated (r > 0.95) for all examined organs. The strong correlations between TLD-measured organ doses and DAP recommend the use of conversion coefficients for the dosimetry of fluoroscopic procedures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.