Abstract
The simultaneous measurement of the spectrally and spatially variant transmittance of a linear variable order-sorting filter in a manner that closely resembles its conditions of actual use is described. The transmittance of a prototype order-sorting filter was measured in the 400- to 880-nm wavelength region by illuminating it with the output beam of a spectrophotometer while the filter was attached to the front of a 30 x 32 pixel silicon array detector. The filter was designed to be used in the output beam of a grating spectrometer to prevent the dispersal of higher diffracted orders onto an array detector. Areas of the filter that were spatially matched to the corresponding detector pixel column had measured peak transmittances of about 90 percent that were uniform to within +/- 1.5 percent along a given column. Transmittances for incident wavelengths shorter than the desired bandpass, corresponding to the order overlap region, were measured in the 0.003 range. Line spread function measurements made with the array detector indicated no significant beam spreading caused by inserting the filter into the beam.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.