Chlorine gas concentrations in the range 0–5 ppm can be measured with a limit of detection of 0.043 ppm using an immobilized reagent ( o-tolidine is best) and refiectance spectrophotometry. The system described utilizes a nylon tape dry reagent carrier whose change in reflectance over 10 s is probed in real time by means of an optical fibre.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.