Abstract
The use of chlorobenzene–ethanol–trimethylpentane solution (CET) in radiation dosimetry is based on radiolytic formation of hydrochloric acid which protonates a pH indicator, thymolsulphonphthalein. The high molar absorptivity of its red form at 552 nm is responsible for a high sensitivity of the system: doses in the range 0.2–15 Gy can be measured. Together with a visual colour comparator it has formed a personnel dosimetry system suitable for accident and civil defense use. A newly constructed optoelectronic reader with two two-colour light emitting diodes (550 and 690 nm) as light sources and a silicone photocell as a detector uses the differential absorbance at these two wavelengths as the response. The response is a direct function of dose and can be recorded and processed electronically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.