Abstract

The purpose of this study was to compare simultaneous measurements of tidal volume (Vt) by optoelectronic plethysmography (OEP) and spirometry during a maximal cycling exercise test to quantify possible differences between methods.Vt measured simultaneously by OEP and spirometry was collected during a maximal exercise test in thirty healthy participants. The two methods were compared by linear regression and Bland–Altman analysis at submaximal and maximal exercise. The average difference between the two methods and the mean percentage discrepancy were calculated.Submaximal exercise (SM) and maximal exercise (M) Vt measured by OEP and spirometry had very good correlation, SM R=0.963 (p<0.001), M R=0.982 (p<0.001) and high degree of common variance, SM R2=0.928, M R2=0.983. Bland–Altman analysis demonstrated that during SM, OEP could measure exercise Vt as much as 0.134L above and −0.025L below that of spirometry. OEP could measure exercise Vt as much as 0.188L above and −0.017L below that of spirometry. The discrepancy between measurements was −2.0±7.2% at SM and −2.4±3.9% at M.In conclusion, Vt measurements at during exercise by OEP and spirometry are closely correlated and the difference between measurements was insignificant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.