Abstract

One of the major medical applications of optoacoustic (OA) tomography is in the diagnostics of early-stage breast cancer. A numerical approach was developed to characterize the following parameters of an OA imaging system: resolution, maximum depth at which the tumor can be detected, and image contrast. The parameters of the 64-element focused array transducer were obtained. The results of numerical modeling were compared with known analytical solutions and further validated by phantom experiments. The OA images of a 3 mm piece of bovine liver immersed in diluted milk at various depths were obtained. Based on the results of modeling, a signal filtering algorithm for OA image contrast enhancement has been proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.