Abstract

In Faraday's magnetometer it is recommended to apply remotely the located poles hemispheres. In this case (and similar cases) there is a need for expeditious and exact positioning of the measuring sensor (Hall) and the studied sample – for obtaining the most authentic characteristics of induction and its gradient, identification and use of a zone of stability. This problem is solved by the corresponding opticalmechanical system of positioning. It consists of the laser modules promoting aim positioning, the Web camera promoting final positioning. On the received concentration dependences of magnetic susceptibility of powder samples existence of limited line sections is confirmed that gives the chance to define susceptibility of particles. The critical relation to use for the similar purposes of samples (disperse) in the form of colloids and suspensions expresses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.