Abstract

To assess the acute results of the first human use of the Optimus covered stent in complex coarctation of the aorta. We successfully implanted the Optimus covered stent in eight cases in patients whose preprocedural anatomy looked challenging for currently available covered stents. Six of the patients had native coarctation with one recoarctation following surgical repair. There were no significant complications with reduction in the mean invasive gradient from 22 to 1 mm Hg. The length of stents used ranged from 33 to 57 mm with a median shortening after expansion of 13%. Postprocedural follow-up with magnetic resonance imaging or computed tomography has not shown evidence of fracture or migration or renarrowing. The median duration of follow-up is 10months. Preliminary results show that the Optimus covered stent is safe and efficacious for use in patients with coarctation of challenging morphology. A systemic trial will be required to evaluate this stent for more widespread practice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.