Abstract

The culture results of 4,807 blood and 383 sterile body fluid specimens received in our laboratory during a 54-month period, were analyzed to determine the time required for culture to become positive, time at which a culture could safely be considered negative, and the spectrum of isolated organisms. The specimens were processed by automated BACTEC 9120 culture system. A total of 1,677 clinically significant microorganisms were isolated. Gram positive and negative bacterial isolation rates were found to be 62.55% and 32.20%, respectively. Yeasts were recovered in 5.24%. False positivity rate was 1.5%. Clinically significant isolates recovered on day four and five were 97.81% and 99.88%, respectively. At day five, the sensitivity was 99.94% and negative predictive value 99.96%. Our data support a five-day incubation protocol for recovery of all clinically significant organisms with sensitivity reduced by 0.06%, when compared with a six-day protocol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.