Abstract
For more effective Tl-201 imaging, the location and width of the energy window set on the Hg-201 x-rays photopeak was investigated using Monte Carlo simulation and phantom experiments. We calculated energy spectra and investigated the amount of primary and scattered photons within various energy windows set on the x-rays photopeak. The energy resolution (ER) at 71 keV (the peak of the x-rays photopeak) was changed to 10%, 12%, 14% and 16%. The relationships between the energy window and the primary counts rate or the scatter fraction (= scattered counts/primary counts, SF) were obtained. By compromise between the primary counts rate and the SF for ER = 12%, the optimum energy window was determined as a wider off-peak window, 77 keV +/- 14.3% (66-88 keV). This off-peak window increased the primary counts rate by 12.5% and decreased the SF by -17% as compared with the conventional on-peak energy window (71 keV +/- 10%, 64-78 keV). When this off-peak widow acquisition was compared with the conventional on-peak window one on a gamma camera, planar and SPECT images using the off-peak widow clearly showed superior results qualitatively and quantitatively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.