Abstract

Alerts generated by intravenous (IV) infusion pump safety software prevent life-threatening situations that might otherwise go unnoticed. However, when alerts are often clinically insignificant, health care workers may become desensitized and discount their importance, resulting in potentially dangerous situations. Little research has been devoted to visual alert desensitization. This paper describes how the Carolinas HealthCare System decreased the number of nonclinically relevant infusion pump alerts by analyzing alert data that were formatted into scatter plots. This in turn enabled the identification of the medications associated with the most meaningful alerts and those associated with the least meaningful alerts. By revising drug library limits for specific medications, it was possible to decrease the number of less clinically meaningful alerts, reduce alert fatigue, and thereby increase the effectiveness of the smart infusion pumps. This added another layer of safety to patient care.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.