Abstract

A method was developed for solid-phase extraction of two chlorotriazine herbicide metabolites, deethylatrazine (DEA) and deisopropylatrazine (DIA), from aqueous samples. Two C 18 phases in cartridge format were compared and recoveries were found to be highly sensitive to sorbent amount, sample volume and presence of parent compounds. Recoveries were significantly improved using a partially non-endcapped C 18 phase compared to the normal C 18 phase, particularly for DIA, apparently due to polar interactions. Combinations of sample volume and sorbent amount were tested using deionized water to determine an optimal combination of 200 ml and 1.0 g, respectively. Recoveries from a variety of river, stream, runoff and ground waters averaged 105–116% and 109–117% at concentrations of 0.5–1.0 ng/ml for DIA and DEA, respectively, with minimum detection limits of 0.05 ng/ml. Other pesticides tested also have acceptable recoveries using this method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.