Abstract

Objective: To compare the sensitivity and specificity of ocular vestibular evoked myogenic potentials (oVEMPs) using 2 electrode montages for the diagnosis of superior canal dehiscence syndrome (SCDS). Subjects: 16 SCDS patients (17 affected-SCDS ears, 15 contralateral-SCDS ears) and 12 controls (24 ears). Methods: oVEMPs were recorded in response to 500-Hz tone bursts using 2 electrode montages. For both montages the active electrode was placed approximately 5 mm below each eye and a ground electrode on the sternum. For montage 1 (standard), the reference electrode was centered 2 cm below each active electrode. For montage 2, the reference electrode was placed on the chin. Results: For either montage, the separation between oVEMP amplitudes in affected-SCDS ears and controls was significant (p < 0.001), with excellent sensitivity and specificity (>90%). Conclusion: oVEMP recordings with the standard montage remain a reliable method for evaluation of SCDS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.