Abstract
Glenoid implant design and alignment are critical to the success of shoulder arthroplasty. Computer-aided design models of 3 glenoid implant designs (keel, standard pegs, and modified pegs) were virtually implanted into 3-dimensional computed tomography-reconstructed models of 40 normal scapulae. The incidence of perforation on 3-dimensional computed tomography preoperative templating varied from 8% to 18% for the 3 models. Malalignment of the implants in retroversion was better tolerated than anteversion and varied with implant design. Tolerance for medialization of the implant to correct for glenoid wear was limited (range, 0 degrees-18 degrees ) and varied with implant design. The modified peg design had the lowest incidence of perforation, the greatest tolerance for malalignment, and the ability to correct for glenoid wear effects. These results emphasize the need for accurate preoperative templating and for developing alignment guides or surgical navigation tools for shoulder arthroplasty and may be used to improve implant fixation designs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.