AbstractWe have operated an atomic force microscope in ambient air with several oscillating cantilever modes to establish the optimal scanning parameters to maximize image resolution and to minimize probe and sample damage. This was done by scanning a surface in air and correlating scan parameters such as oscillation amplitude and damping with image resolution. We also examined the geometry of the probe with a scanning electron microscope, before and after scanning, in order to determine whether the scanning technique had an effect on the geometry of the probe tip. If the probe is oscillated such that it contacts the surface on each oscillation, substantial damage or “wear” to the probe occurs and significant degradation of image quality was observed. In ambient air, the optimal conditions are achieved when the probe penetrates the contamination layer and reverses direction before touching the surface. Under these “near‐contact” conditions no probe damage is observed and high‐image resolution can be maintained indefinitely.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.