Abstract

Microcirculation oxygen levels and blood volumes should be reflected in measurements of myocardial T(2) relaxation. This work describes the optimization of a spiral imaging strategy for robust myocardial T(2) measurement to minimize the standard deviation of T(2) measurement (sigmaT(2)). Theoretical and experimental studies of blurring at muscle/blood interfaces enabled the derivation of parameter sets which reduce sigma T(2) to the level of 5%. T(2) relaxation mapping within healthy volunteers provided estimation of residual sigmaT(2) within the optimized technique. The standard deviation in T(2) measurement across regions of interest (ROIs) in different locations is about 9%. The standard deviation in T(2) measurement in an ROI across different time points is about 5%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.