Abstract

In cochlear implant (CI) users, measurements of electrically evoked compound action potentials (ECAPs) prove the functionality of the neuron-electrode interface. Objective measures, e.g., the ECAP threshold, may serve as a basis for the clinical adjustment of the device for the optimal benefit of the CI user. As for many neural responses, the threshold determination often is based on the subjective assessment of the clinical specialist, whose decision-making process could be aided by autonomous computational algorithms. To that end, we extended the signal-to-noise ratio (SNR) approach for ECAP threshold determination to be applicable for FineGrain (FG) ECAP responses. The new approach takes advantage of two features: the FG stimulation paradigm with its enhanced resolution of recordings, and SNR-based ECAP threshold determination, which allows defining thresholds independently of morphology and with comparably low computational power. Pearson’s correlation coefficient r between the ECAP threshold determined by five experienced evaluators and the threshold determined with the FG-SNR algorithm was in the range of r = 0.78–0.93. Between evaluators, r was in a comparable range of 0.84–0.93. A subset of the parameters of the algorithm was varied to identify the parameters with the highest potential to improve the FG-SNR formalism in the future. The two steps with the strongest influence on the agreement between the threshold estimate of the evaluators and the algorithm were the removal of undesired frequency components (denoising of the response traces) and the exact determination of the two time windows (signal and noise and noise only).”The parameters were linked to the properties of an ECAP response, indicating how to adjust the algorithm for the automatic detection of other neurophysiological responses.

Highlights

  • A cochlear implant (CI) is an auditory prosthesis used to restore hearing in people with severe to profound hearing loss

  • The FG-signal-to-noise ratio (SNR) algorithm and all six evaluators agreed in their classification of evoked compound action potentials (ECAPs) presence or ECAP absence in 64% (100/156) of the recordings

  • In 93 cases (59.6%), all evaluators and the algorithm agreed on the presence of an ECAP response, and in seven cases (4.5%) all agreed on its absence

Read more

Summary

Introduction

A cochlear implant (CI) is an auditory prosthesis used to restore hearing in people with severe to profound hearing loss. Its multi-electrode array is normally inserted into the scala tympani. Automatic ECAP threshold determination and last author, since each of them published the appropriate method previously. The method to implement of the Data Processing Pipeline (DPP) was devised by PS, MT, LG, and SH. Software programming of DPP was realized by MT, KL, and PS. Pseudonymized data were processed and statistically analyzed by MT and KL. Visualization, i.e. creation of figures was done by PS, MT, KL, SH, and LG. The original draft was written by KL, PS, MT, SH, and LG. All authors have read and edited the manuscript

Objectives
Results
Discussion
Conclusion