Abstract

Selective detection of lactate signals in in vivo MR spectroscopy with spectral editing techniques is necessary in situations where strong lipid or signals from other molecules overlap the desired lactate resonance in the spectrum. Several pulse sequences have been proposed for this task. The double-quantum filter SSel-MQC provides very good lipid and water signal suppression in a single scan. As a major drawback, it suffers from significant signal loss due to incomplete refocussing in situations where long evolution periods are required. Here we present a refocused version of the SSel-MQC technique that uses only one additional refocussing pulse and regains the full refocused lactate signal at the end of the sequence.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.